ENGLISH
Conducted RF
3 V
IEC61000-4-6
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
Radiated RF
10 V/m
IEC61000-4-3
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UT is the a.c. mians voltage prior to application of the test level.
Table 3
Guidance and manufacturer's declaration – electromagnetic immunity
Test
Band
Frequency
(MHz)
(MHz)
385
380
–390
Radiated RF
450
430
IEC61000-4-3
–470
(Test
specifications for
enclosure port
710
704 –
immunity to
RF wireless
787
745
communications
780
equipment)
810
800 –
960
870
930
1720
1700
–1990
1845
1970
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Service
Modulation IEC 60601-1-2
Test Level
(V/m)
TETRA 400
Pulse
27
modulation
18 Hz
GMRS 460,
FM
28
FRS 460
±5kHz
deviation
1 kHz sine
LTE Band 13,
Pulse
9
17
modulation
217 Hz
GSM
Pulse
28
800/900,
modulation
TETRA 800,
18 Hz
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
Pulse
28
CDMA 1900;
modulation
GSM 1900;
217 Hz
DECT;
LTE Band 1, 3,
4, 25; UMTS
2450
Radiated RF
IEC61000-4-3
(Test
specifications for
enclosure port
immunity to
5240
RF wireless
5500
communications
5785
equipment)
Table 4
Guidance and manufacturer's declaration - electromagnetic Immunity
Radiated RF
IEC61000-4-39
(Test specifications
Compliance
for enclosure port
level
immunity to
(V/m)
proximity magnetic
fields)
27
Attention: With the exception of energy exchange and cables sold by ma-
nufacturers of lung function devices as spare parts for internal componen-
ts, the use of accessories and cables other than those specified will result
28
in increased product emission or reduced anti-interference.
The following cable types must be used to ensure compliance with inter-
ference radiation and immunity standards.
9
Table: Cable overview
Number
Model
28
1
Power adapter cable
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies
28
Disposal: The product must not be disposed of along with other domestic waste.
The users must dispose of this equipment by bringing it to a specific recycling
point for electric and electronic equipment
2400
Bluetooth,
Pulse
–2570
WLAN,
modulation
802.11 b/g/n,
217 Hz
RFID 2450,
LTE Band 7
5100
WLAN 802.11
Pulse
–5800
a/n
modulation
217 Hz
Test
Modulation
IEC 60601-1-2
Frequency
Test Level
(A/m)
30 kHz
CW
8
134,2 kHz
Pulse
65
modulation
2.1 kHz
13,56 kHz
Pulse
7,5
modulation
50 kHz
Cable length (m)
1.50
26
28
28
9
9
Compliance level
(A/m)
8
65
7,5
Mask or no
Remark
YES
/