21
decide the denote value, after that, it will automatically return to (Data
management) interface.
Figure 18 Denote value setting interface
(5) Exit
Under (Data management), select "Exit" to return to (Menu) interface.
c. Settings
Under (Menu) interface, select "Settings" to enter (Settings) interface as
shown in Fig.19. Under this interface, settings of language, Bluetooth on/
off, time and calibration, and view device information can be realized.
Figure 19 Setting interface
(1) Language setting
Under (Settings) interface, select "Language" to enter (Language setting)
interface as shown in Fig.20. Select "English", the device language will be
English, select "中文", the device language will be Chinese, after selected,
it will automatically return to (Settings) interface.
(2) Bluetooth
Move selection toolbar to "Bluetooth", press "Confirm" key to select "ON"
or "OFF" that can turn on or off the Bluetooth module (If there is no Blue-
tooth module in the device, the operation is invalid).
(3) Time setting
Under (Settings) interface, select "Time" to enter (Time setting) interface
Figure 20 Language setting interface
as shown in Fig.21. Select "Minute" to enter (Minute setting) interface,
as shown in Fig.22. Press "Up" or "Down" key to change the value (long
pressing is available), then press "Confirm" key to return to (Time setting)
interface.
The operation of "Hour", "Day", "Month", "Year" is similar to the "Mi-
nute". The "Week" will be calculated according to "Year", "Month" and
"Day", which does not need to set manually. Then select "Exit" to return
to (Settings) interface.
Figure 21 Time setting interface
(4) Calibration
Under (Settings) interface, select "Calibration" to enter (Calibration set-
ting) interface as shown in Fig.23. Select 2L or 3L based on the volume of
syringe, then enter to (Calibrate) interface as shown in Fig.24.
Figure 23 Calibration setting interface
Under (Calibrate) interface, push the syringe once, the device will display
"REPEAT", then push the syringe once again. After twice correct continuo-
us operation, the calibrating will be succeed, and the device will display
"OK!". Finally the interface will jump to the former interface before ca-
libration (The former interface: If the device is calibrated after measure-
ENGLISH
Figure 22 Minute setting interface
Figure 24 Calibrate interface