ENGLISH
ment completed, it will return to (Settings) interface; if calibrated before
measurement completed, it will return to (Testing) interface.).
If the device displays "Error!Please repeat", it indicates something wrong
with the operation, please repeat the calibrating until succeeded. If the
device displays "Select right volume", please confirm whether the volume
of syringe and calibration selection is accordant, then repeat the calibra-
ting until succeeded. If you need to stop calibrating, just press the"Con-
firm" key to exit to the former interface before calibration.
Under (Calibration setting) interface, select "Adjust" to enter (Adjusting)
interface, as shown in Fig.25. Press "Up" or "Down" key to change the
value (long pressing is available), then press "Confirm" key to return to
(Adjusting confirm) interface, as shown in Fig.26. Selecting "Yes" will save
adjusted value, selecting "No" will cancel the setting, then the device will
return to (Calibration setting) interface.
Note: The value determines the accuracy of measurement, please
do NOT change it randomly. After the turbine has been replaced, cali-
bration shall be applied for inputting parameters of new turbine, which
guarantees the accuracy of measurement after turbine replaced.
F igure 25 Adjusting interface
Under (Calibration setting) interface, select "Exit" to return to (Settings)
interface.
(5) About device
Under (Settings) interface, select "About" to enter (About) interface. User
can view device name and software version. Press "Confirm" key to return
to (Settings) interface.
(6) Exit
Under (Settings) interface, select "Exit" to return to (Menu) interface.
d. Power off
Under (Menu) interface, select "Power off", the device will shut down.
Figure 26 Adjusting confirm interface
Note: If there is no operation within 1 minute, the device will power off
automatically.
e. Exit
Under (Menu) interface, select "Exit" to return to (Main interface). If the
measurement is not completed before enter (Main interface), it will return
to (Testing) interface.
6.1.5 Repeated measure
Measurement of the device is repeatable. Long press "Repeated measure"
key to enter (Testing) interface. When the memory is full, it will display
(Memory full) interface as shown in Fig.27. If you select "Yes", it will enter
(Delete data) interface; if you select "No", it will enter (Menu) interface.
6.1.6 Charge
There are two kinds of charging methods:
1) Connect the device with computer by data line― then the device
should be under charging state.
2) Connect the device with power supply by power adapter, then the devi-
ce should be under charging state.
For device charging, connect it with the power where easy to be cut
off, after charging completed, unplug the power adapter to cut off from
power.
6.1.7 Upload Data
Install the PC software in the computer, then the following figure will ap-
pear after completing.
Figure 27