Technical Information; Failure And Error Messages; Customer Service; Technical Data - uebe cyclotest 2 plus Mode D'emploi

Masquer les pouces Voir aussi pour cyclotest 2 plus:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Technical information

Failure and error messages

Failure en-
Possible cause
countered
Display "L"
- Sensor in the warm-up phase
- Temperature too low
- Possible technical fault
Display "H"
- Temperature too high
- Fever
- Illness
- Possible technical fault
Display "E"
- Measured value not stable
- Sensor removed before end
of measurement

Customer service

The unit may only be repaired by the manufacturer or by an expressly
authorised organisation.
Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Allemagne
Phone: +49 (0) 9342/924040
Fax:
E-mail: info@uebe.com
Internet: www.uebe.com
If you are returning a unit from a non-EU country, please observe your
national customs regulations.
EN-78
Corrective action
- Continue measuring until end of
measurement
- Repeat measurement
- If "L" is permanently displayed,
contact cyclotest® service hotline
- If "H" is permanently displayed
without fever, contact cyclotest®
service hotline
- Repeat measurement
- If "E" is repeatlitly displayed,
contact cyclotest® service hotline.
+49 (0) 9342/924080
.........................
Technical information

Technical data

Unit type:
Cycle temperature computer for recording basal body temper-
ature
Method:
Symptothermal method modified on the basis of the recognised
regulations of the WHO
Unit dimensions:
8 cm x 10.8 cm x 3.3 cm
Weight:
approx. 110 g
Housing material:
PVC-free, ABS-base plastic
Temperature
PVC-free, TPE-base flexible sensor
sensor:
Measurement
Special cable with strain-relief device (kevlar)
cable:
Sensor:
Thermistor
Display:
LCD display (liquid crystal display)
Display elements:
Day, temperature value in °C, baby symbols, battery symbol,
time-axis display, fault indications, direction arrow
Temperature
Absolute in °C (non-extrapolating)
display:
Measurement
35.5 - 38.0 °C
range:
Measurement
±0.1 °C at ambient temperature 10 °C to 35 °C, rel. air humidity
accuracy under op-
up to 93 % non-condensing, air pressure 700 to 1040 hPa
erating conditions:
Storage and trans-
Ambient temperature -10 to +50 °C, Rel. air humidity up to 93
port conditions:
% (non-condensing), Protect against moisture/humidity
Serial number:
The unit has a serial number which is identified with SN: please
quote this number in the event of possible service enquiries.