Problem/
Possible cause/remedy
question
Why should
There are two reasons for this:
the atomiser
1. To ensure a therapeutically
be replaced
e ective particle spectrum, the
regularly?
nozzle hole must not exceed a
specific diameter. Due to me-
chanical and thermal stresses,
the plastic is subject to a certain
degree of wear. The atomiser
insert [9] is particularly sensitive.
This can also alter the compo-
sition of the aerosol droplets,
which directly a ects the e ec-
tiveness of the treatment.
2. Regularly changing the atomiser
is also recommended for hygie-
ne reasons.
Should each
Yes, this is absolutely essential for
person have
hygiene reasons.
their own
atomiser?
11. TECHNICAL SPECIFICA-
TIONS
Type
IH 58
Dimensions
(W x H x D)
110 x 62 x 47 mm
Weight
176 g
Operating
pressure
Approx. 0.25 to 0.5 bar
Atomiser filling
Min. 2 ml
volume
Max. 6 ml
Medication
flow rate
Approx. 0.25 ml/min.
Sound pressure Max. 45 dBA
(acc. to DIN EN 13544-1 section
26)
Mains connec-
Input: 100 – 240 V~; 50 – 60 Hz;
tion
0.5 A
Output: 5 V
Expected ser-
vice life
400 h
Operating con-
Temperature: +10 °C to +40 °C
ditions
Relative humidity: 10% to 95%
Storage and
Temperature: -20 °C to +60 °C
transportation
Relative humidity: 10% to 95%
conditions
Ambient pressure: 700 to 1060
hPa
Aerosol proper-
1) Aerosol delivery: 0.15 ml
ties
2) Aerosol delivery rate: 0.03 ml/
min
3) Particle size (MMAD): 4.12 µm
The serial number is located on the device or in the
battery compartment.
Subject to technical changes.
Particle size diagram
Measurements were performed using a sodium flu-
oride solution with a "Next Generation Impactor"
(NGI).
The diagram may therefore not be applicable to sus-
pensions or highly viscous medications. You can
obtain more detailed information from the manufac-
turer of your medication.