Product Descriptions
The ConformX® overlay design comprises of a 2-way turn visco elastic foam with a 2-way stretch,
vapour permeable, water resistant cover.
Clinical applications
Indications for Use
ConformX overlay is suitable for comfort only.
WARNING
The cover of this product is vapour permeable but
risk. It is the responsibility of the care giver to
ensure that the user can use this product safely.
Instructions for Use
1.
Remove all packaging.
2.
Lay the ConformX mattress overlay on
top of a normal mattress, ensuring that
there are no sharp edges which might rip
the cover. Ensure that the contoured side
of the mattress is uppermost.
3.
Use the four corner straps to safely
secure the overlay to the underlying
mattress.
4.
Ensure the correct month is showing
on the Turntable
TM
of the cover. If the incorrect month is
showing, rotate the mattress to show
correct month at the foot end. Months are
indicated as follows:
1/12 = January
2/12 = February, etc.
Cover
Concealed zip
indicator at the foot
TURNTABLE
Indicator & Cleaning
Instructions
Corner
straps
5.
Use the TURNTABLE
guide for the monthly rotation and turning
of the mattress. This helps to even out
wear of the mattress and to extend the
life of the product.
CAUTION
Do not use the mattress without a cover. It provides
a protective barrier against biological contamination
of the foam.
Decontamination
To Clean the Cover
Wipe all exposed surfaces with a cloth which
has been moistened in simple detergent and
water. Ensure to remove any organic debris
that may be adherent to the surfaces. Dry
thoroughly.
To Disinfect the Cover
For disinfection we recommend a chlorine-
releasing agent such as NaOCI or NaDCC
diluted to 1,000 ppm (range may vary
between 250 & 10,000 ppm depending on
contamination status and local policy).
Wipe with a cloth moistened with clean water
and dry thoroughly.
Alternatively, use a cloth moistened with
alcohol (70%) all allow to air dry.
Routine checks
The foam core and cover should be inspected
regularly for any signs of damage, soiling or
staining.