TROUBLESHOOTING
Before performing any operation, switch off the appliance and disconnect it from the mains.
Problem
Cause
Power supply connections
were not made
Thedevice is not
correctly or in the case of use
working
with battery
the latter is low
The medicine was not inserted
into the nebuliser
The nebuliser has not been
assembled correctly.
The nebuliser is obstructed
The device does not
nebulise or nebulises
poorly
The accessories are not
properly connected to the
appliance
The pipe is bent or damaged
or twisted
The air filter is dirty
The device is louder
The filter is inserted
than usual
If, after checking the conditions described above, the device still is not working properly, we recom-
mend that you contact your trusted dealer or an authorised FLAEM service centre nearest to you. You
can find a list of all Service Centres at http://www.flaemnuova.it/it/info/assistenza
Remedy
Make the power connections correctly and in the
case of a low battery proceed to recharging
Pour the medicine into the nebuliser
Disassemble and reassemble the nebuliser
correctly according to the connection diagram on
the cover.
Hygienic preparation of the nebuliser. Deposits
of medicine due to lack of hygienic preparation
of the nebuliser impair its efficiency and function.
Strictly follow the instructions in the HYGIENIC
PREPARATION chapter.
Check the correct connection between the air
inlet of the appliance and the accessories (see
connection diagram on cover).
Unwind the pipe and check it for crushing or
punctures.
Replace it if necessary.
Replace the filter
Insert the filter properly into the housing
21

Publicité

loading