Variation in the operating range
Distortion magnitude
Ambient
temperature
Humidity
Power supply
Frequency of field
Level of field
Interchangeability
of sensors
All distortion
magnitudes
Adherence to standards
Class III instrument IEC 1010.
Electrostatic discharge (IEC 801-2 and IEC 1000-4-2)
Class of severity :
- level 2 (4 kV) no destruction of build components, but change of function that can be reset
by a new instruction.
- level 4 (15 kV) non destructive.
Radiated electrical fields (EN 55081-2 class B)
Protection from electromagnetic fields in accordance with standard EN 55082-2 up to 200V/m.
Class of severity: level 4 (200 V/m)
Limit of range
0 to 50°C
10 to 90%
no condensation
7.5 to 11 V
20 to 500 MHz
1 MHz to 1 GHz
1 MHz to 2.5 GHz
0.1 to 10 V/m
0.1 to 100 V/m
0.1 to 200 V/m
Frequency from
1 MHz to 2.5GHz
Level of field
from 0.1 to 200V/m
from 0 to 50°C
from 10 to 90% RH
power supply from
7.5 to 10V
frequency from
1 MHz to 2.5GHz
level from
0.1 to 200V/m
Magnitude distorted
all magnitudes
all
magnitudes
all magnitudes
all magnitudes
all magnitudes
all magnitudes
all magnitudes
Threshold of
alarm detection
80
MAX variation
0.3 % / °C
of the reading
± 0.5V/m per 10°C
< 0.5 V/m
0.05 %/V
± 1 dB
± 1.5 dB
± 2 dB
± 0.5 V/m
± 1 dB
± 2 dB
± 1dB
± 0.5dB
± 0.2V/m
of the programmed
value