SET UP INSTRUCTIONS (CONTINUED)
pressing ZERO / TARE. Users will see dashes on the screen as the scale is
finding zero.
PU2 = E
Enabling this feature means the scale will zero itself when it is powered on. After
each use, the scale will clear the weight of the patient and set itself back to the
initial zero level it found on start-up. Health o meter® Professional recommends
keeping this setting as Enable.
2. Choose the setting you prefer and press CLEAR/REWEIGH to save your selection and exit
the Options mode.
3. The scale will restart. The display will show "UEr", followed by a version number, and then
dashes. When the display shows "0.0" the scale is ready to use.
®
Everlock
(500KL only, this feature is not available on 500KG)
®
The Everlock
feature allows the measuring unit to be permanently locked on the scale, disabling
the UNIT button. The Everlock
®
Everlock
.
Warning: This lock is permanent and cannot be reversed.
Notes:
• Press CLEAR/REWEIGH at any time to exit the Everlock® procedure.
• Once Everlock
®
is engaged the Unt and UdP options will no longer be displayed when
scrolling through the Options menu.
• There is no method to undo Everlock
1. While the scale is turned off, press and hold the ENTER button then press and release the
button. Hold the ENTER button until "AOF=d" or "AOF=E" appears on the display. Use
the ▲ or ▼ buttons to advance the display screen until "ELC=d" appears. This is the default
setting with Everlock
2. "PASS1" will appear on the display. This prompts the user to enter a password. Press the
following keys in this sequence:
BMI, ENTER, ▼, ▲
3. If entered correctly, "ELC=L" (if scale was in LB mode) or "ELC=6" (if scale was in KG mode)
will appear on the display.
4. Use the ▼ and ▲ keys to toggle between (L) Pounds and (6) Kilograms. When the desired
unit is displayed, press ENTER and "PASS2" will appear on the display.
5. This will prompt the user to enter the second password. Press the following keys in this
sequence:
▲, ▼, ENTER, BMI
6. If entered correctly, the scale will display "ELC = E".
7. Press CLEAR/REWEIGH to save your selection and exit the Options mode.
8. The scale will restart. The display will show "UEr", followed by a version number, and then
dashes. When the display shows "0.0" the scale is ready to use.
Warning: Do not power off the scale until CLEAR/REWEIGH is pressed to save the Everlock
option, the scale has restarted and the display has returned to "0.0".
P/N UM500KL 500KG Rev 20191120
default setting is "disabled". Follow this procedure to engage
®
®
without sending the scale back to the manufacturer.
®
disabled. Press the ENTER button to start the Everlock procedure.
(Warning: Entering this password will engage Everlock)
273
Back to Home Page