Instruction and Declaration of Manufacturer – Electromagnetic Immunity
The BTL-4000 Smart/Premium is intended for use in the electromagnetic environment as specified below. The
customer or user of the BTL-4000 Smart/Premium device shall ensure that it is used in such environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
NOTE: U
is the AC MAINS VOLTAGE before the application of the test level.
T
BTL-4000 SMART/PREMIUM – MODE  D'EMPLOI I PAGE 54 DE 57
Test level according to
IEC601-1-2
±6 kV for contact
±8  kV  for  air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
<5 % U
(>95 % dip) for
T
0.5 cycle;
40 % U
(60 % dip) for 5
T
cycles;
70 % U
(30 % dip) for 25
T
cycles;
<5 % U
(>95 % dip) for 5
T
sec.
3 A/m
JFB MEDICAL
46 route des bancs de pierre
69690 Bessenay
T 04 74 70 89 08
Email jfb@jfb.fr
Site www.jfb.fr
Electromagnetic environment –
Compliance
level
instruction
±8  kV  air
The floors shall be wood, concrete or
of ceramic tiles. If the floors are
covered with synthetic material, the
relative humidity shall be at least 30
%.
±2 kV power
The supply mains quality shall be that
supply
of a typical commercial and/or hospital
environment.
±1 kV
input/output
±1 kV d. m.
The supply mains quality shall be that
of a typical commercial and/or hospital
environment.
±2 kV c. m.
<5 % U
0.5 c.
The supply mains quality is that of a
T
typical commercial and/or hospital
environment. If the user of the BTL-
40 % U
5 c.
T
4000 Smart/Premium requires
permanent operation during power
mains interruptions, it is
70 % U
25 c.
T
recommended that the BTL-4000
Smart/Premium is powered from an
<5 % U
5 sec.
uninterruptible power supply or
T
battery.
3 A/m
Power frequency magnetic fields shall
be at the levels characteristic for a
typical location in a typical commercial
and/or hospital environment.