impulsions combinées
8.2.2.11
Type
Intensity - CC mode
Intensity - CV mode
Pulse
Frequency
Modulation
impulsions de stimulation (pour la stimulation selon électrodiagnostics)
8.2.2.12
Type
Intensity - CC mode
Intensity - CV mode
Pulse
Pause
Sound of the pulse generation
Impulsions trapézoïdales
8.2.2.13
Type
Intensity - CC mode
Intensity - CV mode
Rise
Pulse
Fall
Frequency (monophasic)
Frequency (symmetric, alternating)
Modulation
8.2.2.14
Impulsions interrompus
Type
Intensity - CC mode
Intensity - CV mode
Frequency of interruption
JFB MEDICAL 46 route des bancs de pierre 69690 Bessenay
T 04 74 70 89 08
BTL-4000 SMART/PREMIUM – MODE D'EMPLOI I PAGE 42 DE 57
asymmetric
0 to 140 mA
0 to 100 V
0.2 to 1,000 ms
0.1 to 550 Hz
See Current Modulation below in this chapter.
rectangular monophasic, triangular monophasic
0 to 140 mA
0 to 100 V
0.1 to 1,000 ms
0.5 to 10 s
beeping according to the pulse length
monophasic, symmetric, alternating
0 to 140 mA
0 to 100V
0 to 250 ms
0 to 250 ms
0 to 250 ms
0.1 to 900 Hz
0.1 to 450 Hz
See Current Modulation below in this chapter.
rectangular, triangular (monophasic, symmetric, alternating)
0 to 140 mA
0 to 100 V
8,000 Hz