* The parameters are defined at the basic pulse frequency 50 Hz.
8.2.2.20
Courant galvanique
Type
Intensity - CC mode
Intensity - CV mode
Stimulation mode
8.2.2.21
Microcourants
Type
Intensity in the CC mode
Pulse
Frequency
Modulation
Note:
8.2.2.22
Stimulations spastique - Hufschmidt
Intensity - CC mode
Intensity - CV mode
Pulses
Delay between channels
Frequency
8.2.2.23
Stimulations spastique - Jantsch
Intensity - CC mode
Intensity - CV mode
Pulses
Delay between channels
Frequency
JFB MEDICAL 46 route des bancs de pierre 69690 Bessenay
T 04 74 70 89 08
BTL-4000 SMART/PREMIUM – MODE  D'EMPLOI I PAGE 45 DE 57
LP-ISO type parameters*: alternation of DF and MF with the amplitude 80 % of DF
continuous, interrupted 8,000 Hz
0 to 80 mA
0 to 100 V
continuous
interrupted (stimulation 1 to 60 s, pause 1 to 60 s)
rectangular, triangular, exponential (monophasic, symmetric,
alternating) and combined
0  to  1,000  µA
0.2 to 1,000 ms (rectangular, combined)
1 to 1,000 ms (other)
0.1 to 1,000 Hz (rectangular)
0.1 to 700 Hz (combined)
0.1 to 900 Hz (other, monophasic)
0.1 to 450 Hz (other, symmetric and alternating)
See Current Modulation below in this chapter.
CC mode only
0 to 140 mA
0 to 100 V
0.1 to 1,000 ms
10 to 3,000 ms
0.15 to 50 Hz
0 to 140 mA
0 to 100 V
0.1 to 1,000 ms
0 to 3,000 ms
0.15 to 50 Hz