product is needed only when the product is removable in the device. The
products are not allowed to contact each other.
2. Use a suitable rinsing adaptor, and connect the internal water lines to
the rinsing connection of the washer-disinfector.
3. Start the program.
4. After the program is finished, remove the product from the washer-
disinfector,
inspect (refer to section "Inspection and Maintenance") and packaging
(refer to chapter "Packaging"). Dry the product repeatedly if necessary
(refer to section "Drying").
Notes
a) Before use, you must carefully read the operating instructions
provided by the equipment manufacturer to familiarize yourself with the
disinfection process and precautions.
b) With this equipment, cleaning, disinfection and drying will be carried
out together.
c) Cleaning: (c1) The cleaning procedure should be suitable for the
product to be treated. The flushing period should be sufficient (5-10
minutes). (c2) Pre-wash for 3minutes, then wash the dirt for another
5 minutes (at least) with pure water at room temperature of 25 °C (or
distilled water/deionised water). (c3) Clean with mild alkaline cleaner at
the condition, concentration and time recommended by the manufacture
of the cleaning agent. Only freshly prepared solutions can be used. (The
used cleaning agent for the cleaning validation was
9