SYMPTOM
Cannotadjust/changetheflowrate.
Yellow Low Oxygen light is illuminated.
OR
Red Service Required light is illuminated with
audible alert.
External Power and/or Battery Power lights
illuminated. Unit is operating.
Red Service Required light is illuminated.
External Power and/or Battery Power lights
illuminated. Audible alert is sounding. Unit
does not operate.
Power Fail Alert activates (External Power
lightflashesandaudiblealertsounds).
RedServicerequiredlightisflashing.Flow
Indicatorlightisflashing.AudibleAlertis
sounding. Unit is operating.
Unit does not operate when using with
charged battery. Power Fail alarm activates.
DC Adapter is attached. Unit runs from
Battery or does not run. The External Power
Present Light not illuminated.
Lowest Battery Status Light is Yellow. Unit
beeps once.
Battery Status Lights never indicate fully
charged.
YellowBatteryStatuslightisflashing.
Audible Alert sounds intermittently when
operating from DC Adapter.
Short Battery Run Time.
If any other problems occur with your iGo.
CARING FOR YOUR DEVILBISS IGO PORTABLE OXYGEN SYSTEM
WARNING
Before attempting any cleaning procedure, turn the unit Off and disconnect from AC or DC power.
Do not use lubricants, oils or grease.
Air Filter
Theairfiltershouldbecleanedatleastonceaweek.Toclean,followthesesteps:
1. Removetheairfilter.
2.
Wash in a solution of warm water and dishwashing detergent.
3. Rinsethoroughlywithwarmtapwaterandtoweldry.Thefiltershouldbecompletelydrybeforereinstalling.
CAUTION–Topreventproductdamage,donotattempttooperatetheunitwithouttheairfilterorwhilethefilterisstilldamp.
NOTE–Devicehasaninternalintakefilterthatmustbeinspectedbyaqualifiedprovider/technicianevery3years.
AC Adapter and Power Cord/DC Adapter
The adapters and power cords should be cleaned as needed. To clean, follow these steps:
1.
Disconnect cords from the iGo before cleaning.
2.
Clean the cords by using a damp cloth with a mild household cleaner and wipe dry.
EN - 10
POSSIBLE CAUSE
1. Flow rate has been locked by your provider.
2. Unit malfunction.
1.Airfilterisblocked.
2. Exhaust is blocked.
1.Airfilterisblocked.
2. Exhaust is blocked.
3. Unit is overheated.
1. Battery completely discharged.
2. Lost external power without battery installed.
1. Blocked or defective cannula or oxygen tubing
1. Battery is overheating.
1. Poor connection to DC power source.
2. Vehicle power source (vehicle accessory connector)
dropped too low for the DC Adapter.
3. Fuse in vehicle is blown.
1. Battery needs to be charged.
1. Battery needs to be conditioned.
2. Defective battery.
1. Unit is equipped with built-in battery test, and the
battery pack is being checked.
1. Vehicle not running.
2. Poor connection to DC power source.
3. Vehicle electrical system overloaded or defective.
1. Battery needs to be reconditioned.
2.Airfilterisblocked.
3. Exhaust is blocked.
4.Intakefiltersilencerisblocked.
REMEDY
1. Contact your DeVilbiss provider.
2. Contact your DeVilbiss provider.
1.Checktheairfilter.Ifthefilterisdirty,washitfollowingthecleaning
instructions.
2. Check the exhaust area and make sure there is nothing restricting the unit
exhaust.
1.Checktheairfilter.If thefilterisdirty,washitfollowingthecleaning
instructions.
2. Check the exhaust area and make sure there is nothing restricting the unit
exhaust.
3a. Allow unit to cool and try again.
3b. Move unit to cooler location.
1. Recharge battery.
2. Install battery or plug into external power.
1.Detachcannula.Ifproperflowisrestored,cleanorreplaceifnecessary.
Disconnecttheoxygentubingattheoxygenoutlet.Ifproperflowisrestored,
check oxygen tubing for obstructions or kinks. Replace if necessary.
1. Allow battery to cool.
1. Ensure the DC accessory outlet is clean and a good connection can be
made. Insert the other end of the DC Adapter into the vehicle's DC accessory
outlet.
2. If the vehicle power source drops too low for the DC Adapter, the iGo will
revert to Battery operation (if present) or will cease operation until power is
restored.
3. Check the fuse and replace if necessary.
1. Plug unit into AC power or replace discharged battery with charged battery.
2. Plug unit into DC accessory outlet to operate unit (battery will not recharge
on DC power).
1. Fully discharge battery then recharge.
2. Contact your DeVilbiss provider.
1.Iflightflasheslongerthan5minutes,batteryisdefective.Contactyour
DeVilbiss provider for a replacement.
1. Start vehicle.
2. Ensure the DC accessory power port outlet is clean, and a good connection
can be made. Insert the DC Adapter into the vehicle's DC accessory outlet.
3.HaveaqualifiedautotechniciancheckelectricalsystemwhileiGois
attached.
1. Perform a complete discharge and then a 100% charge of the battery.
2.Checktheairfilter.Ifthefilterisdirty,washitfollowingthecleaning
instructions.
3. Check the exhaust area and make sure there is nothing restricting the unit
exhaust.
4. R eplacetheintakesilencer/filter.
1. Turn your unit Off and switch to your reserve oxygen system. Contact your
DeVilbiss provider immediately.
A-306-1