  • FRANÇAIS, page 17
Safety distance to the
11 cm
eyes
IP protection class
IP 21
Expected service life of
Information on the service life of the
the device
product can be found on the homepage
This equipment complies with European standard EN 60601-1-
2 and is subject to specific safety provisions for electromagnetic
compatibility. Please note, however, that portable and mobile HF
communication equipment may influence this unit. More details can
be requested from the stated Customer Service address or found at
the end of the instructions for use.
This device meets the requirements of the European Medical De-
vices Directive 93/42/EEC with amendments as well as the German
Medical Devices Act.
The serial number is located on the device or in the battery com-
partment.
If the device is not used according to the specifications, perfect
functionality cannot be guaranteed! We reserve the right to make
technical changes to improve and develop the product.
This device conforms with the European standard EN 60601-1-2
(Group 1, Class B, in accordance with CISPR 11, IEC 61000-3-2,
IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4,
IEC 61000-4-5, IEC 61000-4-6, IEC 61000-4-11, IEC 61000-4-39)
and is subject to particular precautions with regard to electromag-
netic compatibility. Please note that portable and mobile HF com-
munication systems may interfere with this device. For more details,
please contact our Customer Services at the address indicated.
Notes on electromagnetic compatibility
• The device is suitable for use in all environments listed in these
instructions for use, including domestic environments.
• The device may not be fully usable in the presence of electro-
magnetic disturbances. This could result in issues such as error
messages or the failure of the display/device.
• Avoid using this device directly next to other devices or stacked
on top of other devices, as this could lead to faulty operation. If,
however, it is necessary to use the device in the manner stated,
this device as well as the other devices must be monitored to
ensure they are working properly.
• The use of replacement parts other than those specified or pro-
vided by the manufacturer of this device can lead to an increase
in electromagnetic emissions or a decrease in the device's re-
sistance to electromagnetic interference; this can result in faulty
operation.
• Keep portable RF communication devices (including peripheral
equipment, such as antenna cables or external antennas) at least
30 cm away from all device parts, including all cables included
in delivery.
• Failure to comply with the above could impair the performance
of the device.
12.2 Protective eyewear
Model
EP 02
P r o t e c t i o n
4-2.5 BEU 1 S - 166 S CE according to EN 171:2002
level
Protective
4-2.5 BEU 1 S - 166 S CE
eyewear
The code for the lenses is comprised of the
designation
following:
- Initial number for IR protection filter (1st part of
code)
- Protection level number (2nd part of code)
- Manufacturer's marking (3rd part of code)
- 1 describes the optical class of the glasses (4th
part of code)
- S describes increased strength (5th part of code)
- 166 is the normative basis (6th part of code)
- CE mark (7th part of code)
Max. 29.1%
Light transmis-
sion level
Min. 17.8%
Maximum av-
780 - 1400 nm: 2.9%
erage spectral
780 - 2000 nm: 32.3%
transmittance
value with
infrared
Notified body:
ECS European Certification Service GmbH
Hüttfeldstraße 50
73430 Aalen, Germany
Identification number: 1883
EU declaration of conformity: https://www.beurer.com/web/we-
landingpages/de/cedeclarationofconformity.php
13. WARRANTY
Further information on the warranty and warranty conditions can be
found in the warranty leaflet supplied.
Warranty restriction: The enclosed infrared lamp – as with all light
sources – is not subject to warranty.
16