• Risk of burns from hot surfaces.
• Allow the device to cool fully before starting repair work.
• Infrared light sources are excluded from the warranty.
Remove the old infrared lamp
1. Place the cooled device with the ceramic glass pane on a flat
surface.
2. Unscrew the 8 screws on the rear panel of the device using a suit-
able Phillips screwdriver (see Figure
3. Set the housing upright. However, allow the front part with the
ceramic glass pane to rest on a raised surface, e.g. a book (see
Figure
).
C
4. Hold the cooled infrared lamp at the right end and carefully press
it to the lef.
5. Now pull the right end of the infrared lamp forwards out of the
socket.
Insert new infrared lamp
NOTICE
Only insert the new infrared lamp using a cloth. Do not touch the
infrared lamp with bare fingers, as the layer of grease on the skin
can damage it.
1. Insert the new infrared lamp into the socket using a cloth. In-
stallation of the new infrared lamp is carried out in reverse order
to removal.
2. Close the device again. Closing the device is carried out in reverse
order to opening.
3. Ensure that no objects remain in the housing and no cables are
trapped.
If you have problems replacing the infrared lamp, please contact
Customer Services.
8.3 Storage
1. Remove the mains plug and allow the device to cool fully before
moving it.
2. If you do not plan to use the protective eyewear for an extended
period of time, we recommend storing it in its original packaging
in a dry environment, ensuring that it is not weighed down by
other items.
3. The device should ideally be stored in the original packaging and
in a dry place.
9. REPLACEMENT PARTS AND WEARING
PARTS
To purchase wear parts and replacement parts, please visit www.
beurer.com or contact the corresponding service address (accord-
ing to the service address list) in your country.
Replacement lamp/type:
Halogen light source R7S
Replacement part:
Protective eyewear
).
B
Technical specifi-
Article
cations:
number:
220-230 V / 300 W
162.620
Article
number:
200.345
10. WHAT IF THERE ARE PROBLEMS?
Problem
Possible cause
Infrared lamp
ON/OFF switch
does not
was not pressed
illuminate
No power
Infrared lamp burn
time exceeded
After replacing the
infrared lamp, the new
infrared lamp was not
inserted correctly
Infrared lamp
Safety switch-o
shuts o dur-
due to overheating
ing operation
protection
although
timer has not
yet elapsed
11. DISPOSAL
For environmental reasons, do not dispose of the device in house-
hold waste at the end of its service life. Dispose of the device at a
suitable local collection or recycling point in your country.
Observe the local regulations for material disposal. Dis-
pose of the device in accordance with EC Directive Waste
Electrical and Electronic Equipment (WEEE). If you have
any questions, please contact the local authorities re-
sponsible for waste disposal. You can obtain the location
of collection points for old devices from the local or municipal au-
thorities, local waste disposal companies or your retailer, for exam-
ple.
12. TECHNICAL SPECIFICATIONS
12.1 Infrared heat lamp
Mains connection
Power consumption
Risk group
Timer
Max. irradiation
intensity according to
EN 62471
Dimensions (W x D
x D)
Weight
Operating conditions
Storage and
transportation
conditions
Emitted infrared
wavelength range
Protection class
Expected burn time of
light source
Fuse
15
Solution
Press the ON/OFF
3
button
3
Plug the mains plug into
the socket correctly
The infrared lamp is
faulty. Contact Cus-
tomer Services or your
retailer
Check that the new
infrared lamp has been
fitted correctly as de-
scribed in Chapter 8.2
Allow the device to cool
down and restart it.
AC 220 V - 240 V / 50-60 Hz
300 W
RG2, medium risk group
Adjustable from 1 to 15 minutes in
minute increments
<3200 W/m
2
270 x 195 x 285 mm
approx. 1,4 kg
Temperature: 10 °C to 35 °C
Relative humidity: ≤ 90 %,
Ambient pressure: 700 hPA - 1060 hPa
Temperature: -10 °C to 50 °C
Relative humidity: ≤ 90 %
Ambient pressure: 700 hPA - 1060 hPa
IRA + IRB / 780-3000 nm
II
2.000 hours
Typ T2A/AC250V

Publicité

loading