13 Environmental Considerations
Product
CPRmeter
CPRmeter Patient Adhesive
Information
The CPRmeter contains electronic
components. Dispose of it at an
appropriate recycling facility in
accordance with local regulations.
The used patient adhesive may be
contaminated with body tissue,
fluid, or blood. Dispose of it as
infectious waste.
28
14 Electromagnetic Conformity
Guidance and manufacturer's declaration: The CPRmeter is intended for use in the
electromagnetic environment specified in the tables below. The user of the CPRmeter
should assure that it is used in such an environment.
Electromagnetic Emmissions
Emissions Test
Compliance
Electromagnetic environment - Guidance
RF CISPR 11
Group 1 Class
The CPRmeter uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The CPRmeter is suitable for use in
all establishments, including domestic
establishments and those directly connected
to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Electromagnetic Immunity
Compliance
Immunity Test
IEC 60601
Test Level
Level
Electrostatic
±6 kV contact
±6 kV contact
Discharge
±8 kV air
±8 kV air
(ESD)
IEC 61000-4-2
Power
3 A/m
3 A/m
Frequency
(50/60/400 Hz)
Magnetic Field
IEC 61000-4-8
Radiated RF
10 V/m
10 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
29
Electromagnetic
Environment -
Guidance
There are no special
requirements with
respect to electrostatic
discharge.
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial/
hospital environment.
There are no special
requirements for
non-commercial/
non-hospital
environments.
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the CPRmeter,
than is absolutely
necessary.†,‡ The
recommended
separation distances for
various transmitters and
the CPRmeter are
shown in the following
table.
Interference may
occur in the vicinity of
equipment marked
with the following
symbol: