Reusable Silicone Cover Kit
The CPRmeter can be fitted with the accessory Reusable Silicone Cover and
the large patient adhesive (single-use). This increases the contact area to the
patient.
Applying the Reusable Silicone Cover and large patient adhesive
1. Check that the silicone cover has been cleaned
and disinfected after last use.
2. Insert the CPRmeter
into the sleeve.
3. Check the label
on the CPRmeter large
patient adhesives resealable
pouch to be sure that they are within their expiration date.
4. Open the package of CPRmeter large patient adhesives and peel one
from the white liner strip to expose the adhesive surface on its underside.
5. Align the bottom of the patient adhesive with the yellow
patient adhesive mount area on the Rear Cover of the
CPRmeter and the adjacent Silicone Cover's bottom surface.
Make sure the channel on the adhesive is directly over the vent
membrane. Press the large patient adhesive into place.
6. Do not peel off the green liner from the attached large patient adhesive
until you are ready to apply the CPRmeter to a patient in an emergency.
Return remaining patient adhesives to their resealable pouch.
IMPORTANT NOTE: Do not delay CPR. If the Reusable Silicone Cover is
generally used with the CPRmeter, the cover and the Large Patient
Adhesives should be applied to the CPRmeter in advance.
Use and maintenance
The CPRmeter use or maintenance is not changed by use of the Silicone
Cover. Clean the Silicone Cover separately from the CPRmeter. Disinfect the
Silicone Cover using the following method:
1. Clean and scrub the cover with a small soft brush, using a mild detergent
until the cover is visibly clean.
2. Rinse in lukewarm water.
3. Submerge the cover in a 0.55% solution of ortho-phthalaldehyde in
accordance with manufacturer's specifications.
4. Rinse again and dry.
The Silicone Cover can also be autoclaved in distilled water at 136 ºC
(210 ºF) and 2.0 kg/cm² for 10-20 minutes.
WARNING: The CPRmeter with the Silicone Cover can not be
considered as disinfected or sterilized.
22
9 Glossary of Symbols
The following symbols appear on the CPRmeter, accessories and packaging:
Symbol
Definition
The product is in compliance with the essential requirements of Council
Directive 93/42/EEC as amended by Council Directive 2007/47/EC.
Compliance with applicable U.S and Canadian safety standards has
been certified by Canadian Standards Association.
These CPRmeter patient adhesives are disposable and are for single
patient use only. Do not re-use. Re-use will lead to increased risk of
cross contamination, and/or degradation of adhesive performance.
Defibrillation protection. The CPRmeter is defibrillation protected,
type BF patient connection.
Manufacturer.
Dispose of in accordance with your country's requirements.
Reference order number.
IP55
The CPRmeter meets IEC 60529 class IP55.
Serial number.
Expiration date for patient adhesives, formatted YYYY-MM
(year-month).
LATEX
The patient adhesives, CPRmeter or other parts do not contain
natural rubber latex.
Warning / Caution symbol
Store the CPRmeter patient adhesives between the temperatures
shown. See Chapter 12, Specifications.
= #
Contains number of CPRmeter patient adhesives shown as "#. "
Refer to Directions for use.
Consult Directions for Use
Lift here to peel off the patient adhesive liner and apply to patient's
bare chest.
Not for use on children under 8 years old
microSD memory card
SD
Battery – 3V Lithium 123 type
Federal law (US) restricts this device to sale by or on the order of
a physician
23