ADDITIONAL OPERATING MODES AND FEATURES
The Display Menu—With the blower on, press the Select button on the keypad to scroll through the items on the display. Not all of these will display
on your unit. Some are optional, depending on what your physician and/or home medical equipment provider has selected. The table below shows the
items you may see and describes the feature associated with each item. Please note that the Displayed Text are examples only.
Displayed
Item
Text
Examples
Actual Pressure
Act Pres
20.0 cmH
Compliance Bar Graph Compl %
70
SpO2/Pulse Rate
SpO2 95
PR 80
CPAP Pressure
CPAP
20.0 cmH
Upper Pressure Limit
Up Limit
(UPL)
15.0 cmH
Lower Pressure Limit
Lo Limit
(LPL)
5.0 cmH
AutoAdjust Delay Time Delay
10 min
CPAP Delay Time
Delay
10 min
CPAP Delay Pressure
Delay Pr
4.0 cmH
Enable Mask Alert
Msk Alrt
Enabled
Humidifier Present
9100 Hum
Attached
Compliance Meter
ComplHrs
123456.7
Hour Meter
Unit Hrs
123456.7
Firmware Version
Firmware
M5 V0.17
Serial Number
Serial #
GD000001
Language Selection
Language
English
SE-9054
A d d i t i o n a l O p e r a t i n g M o d e s a n d F e a t u r e s
Description
Actual mask pressure calculated from
pressure and flow.
The bar graph is displayed to give feedback on
device usage. Graph covers usage over the last
30 days of operation (rolling 30-day basis).
If an oximeter is used, SpO2 and PR values
will display.
Prescription setting – this will be seen only
if your AutoAdjust CPAP is to be used in the
CPAP mode.
Prescription setting – maximum AutoAdjust
pressure (AutoAdjust mode).
Prescription setting – minimum AutoAdjust
pressure (AutoAdjust mode).
Selectable delay time setting.
Selectable delay time setting in CPAP Mode. 0 to 45 minutes
Pressure during the delay in CPAP Mode.
Indicates whether the 20 second audible
mask alert feature is enabled or disabled.
Indicates that a DeVilbiss humidifier is
connected to the CPAP.
The total number of hours the AutoAdjust
CPAP has sensed breathing.
The total number of hours that the
AutoAdjust CPAP blower has been ON.
Shows version of installed firmware.
Shows the serial number of the AutoAdjust
CPAP unit.
Shows the selected LCD display language.
Range
3-20 cmH
O
The value displayed may vary slightly from
2
your prescription based on operating
conditions.
0-100%
The value displayed is your actual use time
versus the time you are expected to use the
AutoAdjust CPAP.
0-100%
displayed for both values if connection
***
18-300 pulse/min
is faulty or probe is not properly placed.
3-20 cmH
O, in
Your provider will set the CPAP prescription
2
0.5 cm
for your AutoAdjust CPAP.
increments
LPL to 20 cmH
O,
Your provider will set the UPL prescription
2
in 0.5 cm
for your AutoAdjust CPAP.
increments
3.0 cmH
O to
Your provider will set the LPL prescription
2
UPL, in 0.5 cm
for your AutoAdjust CPAP.
increments
0-240 minutes in
Your provider has set the delay value.
10 minute
increments
You can select the delay you desire using the
in 5 minute
Up/Down buttons if the provider has
increments
unlocked this setting.
3 cmH
O to CPAP
Your provider has set the delay pressure
2
in 0.5 cm
value.
increments
Enabled or
The mask alert can be enabled or disabled by
Disabled
using the Up/Down buttons unless the
provider has locked the alert.
Attached or
The indication of the attachment or removal
Removed
of the humidifier can be changed using the
Up/Down buttons unless the provider has
locked this setting.
N/A
N/A
N/A
N/A
Up to 15
The display language can be changed using
selections
the Up/Down buttons, unless the provider
has locked the language.