4. Operation

NOTE: Always read the operating instructions before use.

4.1 GENERAL OPERATION

Static
Statique
Alternate
Alterné
Pulsation
Pulsation
Seat Inflation
Gonflage Du Siège
4.2 CPR
Harmony True Low Air Loss with Alternating Pressure System
Système À Faible Vraie Perte D'air Et Pressurisation Alternée
lb
lb
Item / Article #14200
100 lb
200 lb
300 lb
400 lb
45 Kg
90 Kg
136 Kg
181 Kg
AutoFirm
Fermeté
Low Pressure
Pression Faible
Automatique
Power Failure
Perte D'alimentation
System Failure
Soutien Technique
1.
Switch on the main power switch found
on the left side of the Control Unit while
facing the User. Press (
panel to turn on the power.
2. The system will automatically go into "Auto
Firm" mode for a few minutes of inflation.
3. When the initial inflation (Auto Firm pro-
cess) is completed, the system will auto-
matically enter into Static mode
4. Using the Comfort Weight Setting buttons,
adjust according to the weight and height
of the patient, adjust the pressure setting
to the most suitable level without bottom-
ing out.
5. Using the Therapy Mode button choose the
desired therapeutic mode.
NOTE: Seat inflate mode should be used for
ingress and egress at all times to prevent bot-
toming out.
When CPR needs to be performed, click on
the release button on Quick Connector and
quickly detach the connector from the system
to release air.
12