ERR06
Tolerance limit is out of the acceptable range
ERR07
OFFSET is out of the acceptable range
The instrument must be returned to the service centre for checking
ERR08
Reference point in „Relative-Mode" is out of the indicating range
ERR09
no communication with computer or peripheral devices
ERR10
interruption of communication during data transfer
ERR11
Pairing information get lost, remove the connected devices and restart pairing!
ERR12
Bluetooth connection has failed after 30 sec., restart pairing!
Safety information
- Before starting please read this manual carefully and follow the safety information.
- The charging station, charging cable or power supply must not be used if damaged. Please contact in this case
the seller/dealer.
- Protect charging station against high temperature and humidity.
- People with a cardiac pacemaker should consult their physician before using the charging station.
- Do not open the charging station.
- Do not place anything between charging station and the gauge to be loaded.
- Do not place magnetic cards or magnetic storage media near the charging station.
Scope of supply (look at page 7)
• Charging station (B1), • Mico-USB cable (B2)
Preparing to use the charging station
1. Connect the micro-USB connector of the cable (B2) to the micro-USB socket of the charging station (B1).
2. Connect the USB-A connector of the cable (B2) to a power supply* or to the USB port of a PC/laptop*.
3. If properly connected, the LED of the charging station is green.
4. Place the gauge correctly into the recess of the charging station (B1).
5. The charging process is acoustically confirmed and the LED changes to blue.
6. After complete charging the charging process stops automatically and the LED of the charging station (B1)
changes to green.
*Not included in delivery.
Caution:
If there is an acoustic signal from the charging station in short intervals and the LED changes permanently
between green and blue the gauge is placed incorrectly or the power supply of the charging station (via USB
hub) is inadequate.
Error report
Inductive Charging Station
19

Publicité

loading

Produits Connexes pour Mitutoyo BA9382M

Ce manuel est également adapté pour:

209-925209-926209-927209-928209-92963aaa158 ... Afficher tout