Easy Calibration

The calibration mode 'CAL'
Adjustment
The instrument should be set to either the nominal volume (for example
200 µl for a 200 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 38 for procedures.
What to do
1.
Bring up the CAL
mode
2.
Activate the CAL
mode
3.
Confirm CAL mode
4.
Set the volume
5.
Confirm volume
54
TM
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
'off'. 'CAL' blinks.
Press one of the arrow
buttons to activate the CAL
mode. The display changes
from 'off' to 'on'. 'CAL'
continues to blink.
Press the ENTER key.
The display now shows
the set pipetting volume.
'CAL' blinks.
Use the arrow keys (+/-)
to set the volume, which
was previously determined
and tested. 'CAL' blinks.
Press the ENTER key.
The display shows the
tested and corrected
volume. The CAL symbol
is continously displayed to
confirm that an adjustment
has been made.
Adjustment of the Transferpette
Keys to press
1x
>3s
1x
1x
®
electronic