maintenance
The insert must be cleaned regularly, to ensure a good standard of hygiene. Use an acid-
based cleaner. Spray the cleaner over the insert and leave to act for five minutes. Run
a process. If there is severe fouling, we recommend cleaning the insert mechanically
by dismantling it or by removing it and placing it in a solvent bath. Replace the knife
when necessary.
Changing the knife
1. To remove the upper part of the
pouch insert, press the flange on the
quick release couplings downwards.
Lift out backward and upward.
3. Remove the pins that secure the knife.
V1763
V1765
assembly is done in the reverse order.
2. Pull out the cotters on the knife.
4. Grasp the knife and carefully
remove it from the attachments.
Be sure to collect the loose nylon
washers between the knife and the
rotating attachment.
V1764
V1766