Bluetooth
status
®
Bluetooth
symbol AO blinking slowly: Bluetooth
and waiting for connection.
®
Bluetooth
symbol AO not blinking: Bluetooth
established.
®
Bluetooth
symbol AO blinking rapidly: Bluetooth
error.
®
In case of Bluetooth
®
tooth
, wait for a minute, then re-try Bluetooth
Refer to chapter «10. Error Messages» for details.
9. PC-Link functions
This device can be used in conjunction with a personal computer
(PC) running the Microlife Blood Pressure Analyzer+ (BPA+) soft-
ware. The memory data can be transferred to the PC by
connecting the monitor via a cable.
If no download-voucher and cable is included download the BPA+
software from www.microlife.com/software and use a Micro-USB
cable.
10.Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Error
Description Potential cause and remedy
«Err 1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
BK
the measurement.*
«Err 2»
Error signal
During the measurement, error signals
were detected by the cuff, caused for
BT-B
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
Abnormal
An adequate pressure cannot be
cuff pressure
generated in the cuff. A leak may have
BT-C
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
8
®
is activated
®
connection
®
connection
connection error, turn off device Blue-
®
connection.
Error
Description Potential cause and remedy
«Err 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be
displayed. Read through the checklist
for taking a reliable measurement and
then repeat the measurement.*
«Err 6» MAM Mode
There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for taking a reli-
able measurement and then repeat the
measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 299 mmHg) OR the pulse is too
high
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
«LO»
Pulse too low The pulse is too low (less than 40 beats
per minute). Repeat the measure-
ment.*
®
Bluetooth
Bluetooth
symbol AO
the device Bluetooth
blinks rapidly
app on the smartphone. Wait for 1
minute, open the app on the smart-
phone and manually activate Blue-
tooth
tooth
«Err bt» Bluetooth
®
Bluetooth
self check
your local Microlife distributor.
error
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
11. Safety, care, accuracy test and disposal
Safety and protection
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
®
connection error. Turn off
®
and close the
®
on the device, to re-try Blue-
®
connection and data transfer.
®
is malfunctioning. Contact