Assembly

WARNING
• Ensure that the axes of the sensor and pipe are perpen-
dicular to one another (see Fig. 4). Ensure that the thread is
aligned correctly. It will otherwise affect the stability of the
measurement points in pressurised processes.
• The axis through the hole of the sensor tip must line up with
the pipe axis (see Fig. 4)
• Observe the minimum distance of 10 mm between the sen-
sor tip and pipe wall.
• The sensor tip must be covered by the water sample at all
times during the measurement.
• The measurement results will otherwise be incorrect.
IMPORTANT
• If you later require measurement accuracy better than 10 %,
fit a tap near the sensor to take water samples (see section 8
under "Calibration").
A PVC DN 40 fusion joint with the correct external thread is
included. A corresponding DN 40 PP welded joint is available as an
optional accessory (order no. 1023365).
Connect the joint with the sleeve provided in accordance with the
instructions provided by the manufacturer of the pipe materials.
Let the joint harden.
Before installing the sensor, carry out commissioning up to and
including zero point calibration (see section 8).
Insert the sensor into the opening of the media tubing or the tank.
When the media is flowing the bore of the sensor tip must be
positioned in the centre of the tube cross section.
Rotate the sensor so that the bore of the sensor tip is facing in
the direction of the flow (arrow).
Tighten the union nut.