CSA Control Drawing

pH-Measuring Loop
Terminals
Uo, Vsc
1/2, 4, 5, 6
IIC (GRP A, B)
10V
IIB (GRP C)
10V
IIC (GRP D)
10V
Temperature Measuring Loop
Terminals
Uo, Vsc
7, 8
IIC (GRP A, B)
5V
IIB (GRP C)
5V
IIC (GRP D)
5V
Sensor Supply
Terminals
Uo, Vsc
17, 18, 19
IIC (GRP A, B)
10V
IIB (GRP C)
10V
IIC (GRP D)
10V
All combined outputs
Terminals:
Uo, Vsc
1/2, 4, 5, 6, 7, 8,
17, 18, 19
IIC (GRP A, B)
10V
IIB (GRP C)
10V
IIC (GRP D)
10V
100
Io, Isc
Po
Co, Ca
10mA
25mW
3µF
10mA
25mW
9µF
10mA
25mW
24µF
Io, Isc
Po
Co, Ca
3mA
4mW
100µF
3mA
4mW
300µF
3mA
4mW
800µF
Io, Isc
Po
Co, Ca
14mA
35mW
3µF
14mA
35mW
9µF
14mA
35mW
24µF
Io, Isc
Po
Co, Ca
27mA
66mW
3µF
27mA
66mW
9µF
27mA
66mW
24µF
Transmetteur pH 2100 e/2(X)H
Lo, La
250mH
1H
1H
Lo, La
1H
1H
1H
Lo, La
170mH
620mH
1H
Lo, La
44mH
180mH
350mH

Publicité