5.2 Sensor setup menu (continued)
Actual sample
pressure
Filtration probe
heating
Humidity in sensor
Heating drain
Process
Remaining time
Maintenance mode
Maintenance output
mode
Start
Reagent level
Cleaning solution level
Standard solution level
Air filter pads
Membrance
replacement
Electrode replacement
Piston pump
Reagent pump
Cleaning solution pump
Pre-pumping
Pre-pump all
Pre-pump reagent
Pre-pump cleaning
solution
Pre-pump standards
Pre-pump filtration
probe
Pre-pumping sample
Clean filter modules
New filtration probe
module
Pump membrane
Compressor
Cleaning...
Flushing
Reset error
Maintenance (continued)
English
All manuals and user guides at all-guides.com
Only if filtration probe is registered: shows the actual minimum pressure at the filter modules
Only if filtration probe is registered: indication of switching state of sample tubing heating
Only if filtration probe is registered: indication of whether there is moisture in the probe
enclosure
only in 1 or 2 channel mode: shows status of drain heating
Indication of what the instrument is doing.
Indication of the time left for the currently ongoing process
Instrument can be adjusted to service mode (system free of liquids, thermal management
and compressor for filter probe (if installed) active
Value that is output in the service state. Hold = last measured value,
Transfer = Transfer value that is programmed on SC controller
Leave service mode, start measurement
Resets the maintenance counter after reagent change
Resets the maintenance counter after changing the cleaning solution.
Resets the maintenance counter after changing the standards.
Menu-based process for changing the air filter pads, resetting the maintenance counter
Menu-based process for membrane-cap change of the electrode, sets new date in the
maintenance counter and resets electrolyte counter
Menu-based process for electrode change, sets new date in the maintenance counter
Days left until the next pump piston and cylinder replacement (AMTAX piston pump), reset
after replacement of the pump
Number of pump strokes performed by the reagent metering pump, reset after replacement
of the pump
Number of pump strokes performed by the cleaning metering pump, reset after replacement
of the pump
All liquids are pre-pumped in succession.
The reagent is pre-pumped.
The cleaning solution is pre-pumped.
The standards are pre-pumped.
Only if filtration probe is registered: The filtration probe and modules are bled and
pre-pumped.
Only if filtration probe is registered: sample is pumped from the filtration probe for 1 minute
Menu-based process for cleaning the filter modules, automatically resetting the maintenance
counter. With filtration probe.
Only if filtration probe is registered: last filter module replacement.
Only if filtration probe is registered: shows the days left until the next change of pump
membrane (Filter Probe sc sample pump), Rest of the counter.
Only if filtration probe is registered: days left until the replacement of the air compressor.
Reset of the counter.
Trigger an automatic cleaning, then start measurement
Pumps all liquids in succession. Put all tubings that go to reagents, standards and cleaning
solutions into deionised water and start FLUSHING prior to taking instrument out of operation
Reset all error messages
Operation
9