Operation
5.2 Sensor setup menu (continued)
mV zero
mV standard 1
mV standard 2
mV slope
Last calibration
mV active
Process
Remaining time
List of values
Maintenance counter
Operating hours
Reagent level
Cleaning solution level
Standard solution level
Electrode replacement
Membrance
replacement
Air filter pads
Piston pump
Reagent pump
Cleaning solution pump
Filter module status
Clean filter modules
New filtration probe
module
Pump membrane
Compressor
Electrolyte warning
Test/Maintenance
Signals
Process
Remaining time
mV acctive
Cuvette temperature
Enclosure
temperature
Cooling
Heating
Analyzer pressure
Humidity in analyzer
Filter module status
Minimum sample
pressure
Maintenance (continued)
8
All manuals and user guides at all-guides.com
Electrode signal at the zero point
Electrode signal with standard 1
Electrode signal with standard 2
Electrode signal change per decade
Time of the last calibration
Current electrode potential
Information what instrument is currently doing (measurement, calibration etc.)
Remaining time for current process, counting down to zero
List of the last 10 measured values
Counter for reagent and consumables
Indicates the instrument's operating hours.
Displays the current level of the reagent.
Displays the current level of the cleaning solution.
Displays the current level of the standard.
Date of the last electrode change
Last electrode membrane and electrolyte change.
Days left until the next air filter change/clean.
Days left until the next pump piston and cylinder replacement (AMTAX piston pump)
Number of pump strokes performed by the reagent metering pump
Number of pump strokes performed by the cleaning metering pump
Only if filtration probe is registered: indicates the state of the modules.
Only if filtration probe is registered: last filter module cleaning.
Only if filtration probe is registered: last filter module replacement.
Only if filtration probe is registered: date of the last pump membrane replacement (filtration
probe sample pump).
Only if filtration probe is registered: days left until the replacement of the air compressor.
counts down from 90 days. Negative values indicate that electrolyte change is overdue.
Reset by "CHANGE MEMBRANE" or "CHANGE ELECTRODE" process
Maintenance processes
Indication of what the instrument is doing.
Indication of the time left for the currently ongoing process
Current electrode potential (For electrode data ranges refer to
Current measuring cell temperature
Current temperature in the instrument
Current speed of the enclosure fan in %
Current enclosure heating power
Current pressure in the metering system of the valve block in mbar
Indication of whether there is liquid in the collecting tray
Only if filtration probe is registered: shows status of filtration modules (0–100 %)
Only if filtration probe is registered :shows the averaged minimum pressure at the filter
modules
Table 10 on page
62).
English