7.2.3 Warnings (continued)
Warning displayed
Instrument reaction
Continued
Cleaning solution level
measurement
Continued
Standard solution level
measurement
Continued
Electrolyte warning
measurement
Continued
Sample 1 / Sample 2
measurement
English
All manuals and user guides at all-guides.com
Cause
Amount of cleaning
solution has dropped
below warning level
set
Amount of standard
solution has dropped
below warning level
set
Elektrolyte level may
be to low or last
CHANGE MEMBRANE
has been more than
90 days ago.
the amount of sample
is not sufficient
(channel1 / channel2)
This occurs as warning
if Sample detection is
set to Waring
Troubleshooting
Solution
Reset warning
Check cleaning
solution level and
replace if necessary,
On the menu
then reset cleaning
MAINTENANCE/
solution level. The
MAINTENANCE
level is indicated
COUNTER/
mathematically and
CLEANING
can only function
SOLULTION LEVEL
reliably if the counter is
only reset when the
solution is changed
Check standard level
and replace if
necessary, then reset
On the menu
standard level. The
MAINTENANCE/
level is indicated
MAINTENANCE
mathematically and
COUNTER/
can only function
SZTANDARD
reliably if the counter is
SOLULTION LEVEL
only reset when the
solution is changed
Reset warning
Check amount of
manually or automatic
electrolyte, exchange
reset after use of
electrolyte completely
MEMBRANE
if level is low. Use
REPLACEMENT or
process CHANGE
ELCTRODE
MEMBRANE . DO
REPLACEMENT or
NOT add new
CHANGE
electrolyte to old
MEASUREMENT
electrolyte.
RANGE in menue
TEST/MAINTENANCE
Check sample delivery,
make shure that
Automatic reset when
sample line has no
enough sample is
negative pressure,
available or manuell
check tightness of
reset
piston pump, check
overflow and air valve
29