Gima SUPER VEGA Mode D'emploi page 10

Masquer les pouces Voir aussi pour SUPER VEGA:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Guidance and manufacturer's declaration – electromagnetic Emissions
The surgical aspirator SUPER VEGA is intended for use in the electromagnetic environment specified below.
The customer or the user of the surgical aspirator SUPER VEGA should assure that it's used in such an
Emissions test
Irradiated / Conducted
emissions
CISPR11
Irradiated / Conducted
emissions CISPR11
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations / flicker
emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration – electromagnetic Immunity
The surgical aspirator SUPER VEGA is intended for use in the electromagnetic environment specified below.
The customer or the user of the surgical aspirator SUPER VEGA should assure that it's used in such an
Immunity Test
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
Electrical fast transient / burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
Magnetic field
IEC/EN 61000-4-8
Conducted Immunity
IEC/EN 61000-4-6
Irradiated Conducted
IEC/EN 61000-4-3
Nota U
is the value of the power supply voltage
T
(as request by regulation EN 60601-1-2:2001)
environment.
Compliance
Group 1
Class [B]
Class [A]
Complies
(as request by regulation EN 60601-1-2:2001)
environment.
Compliance
± 6kV on contact
± 8kV in air
± 2kV power supply
± 1kV differential mode
5%U
for 0.5 cycle
T
40%U
for 05 cycle
T
70%U
for 25 cycle
T
<5%U
T
for 5 sec
3A/m
3Vrms 150kHz to 80MHz
(for appliances that aren't
life - supporting)
3V/m 80MHz to 2.5 GHz
(for appliances that aren't
life – equipment)
10
Electromagnetic environment -
guidance
The surgical aspirator SUPER VEGA only
used RF energy only for its internal
functioning. Therefore, its RF emissions
are very low and are not cause
interference in proximity of any
Electronic appliances.
The surgical aspirator SUPER VEGA can
be used in all environments, including
domestic and those connected directly to
the public mains distribution that supplies
power to environments used for domestic
scopes.
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial environment or
hospital.
Mains power quality should be that of a
typical commercial environment or
hospital.
Mains power quality should be that of a
typical commercial environment or
hospital. If the user of the surgical
aspirator SUPER VEGA request
that the appliance operates continuously,
the use of a continuity unit is
recommended.
The power frequency magnetic field
should be measured in the intended
installation location to assure that it's
sufficiently low.
-
-

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Vega

Table des Matières