Seca 655 Mode D'emploi page 93

Publicité

Les langues disponibles

Les langues disponibles

Item
Device component
Display interface
12
5.3 Setting up the device
Foot screw
• Used to supply the multifunctional display with power
• Used for data transmission between the multifunctional display and the device
electronics
To achieve accurate measuring results, the floor at the setup location must be
level and stable. Soft floors (wooden boards, for example) give under the
patient's weight and falsify the measuring result.
1. Place the device on a firm, level surface.
You have the following options for continuing:
► Device with ultrasonic measuring rod: Continue with step 2.
► Devices without ultrasonic measuring rod: Continue with step 3.
2. Mark the area shown in the illustration, e.g. using colored adhesive tape.
3. Align the device by turning the foot screws.
The device is positioned horizontally when the air bubble of the spirit level
is precisely in the middle of the circle.
CAUTION!
Injury from a lack of stability
If the foot screws are screwed out too far, they can come loose from
the device. The device is then unstable.
► Screw the foot screws out a maximum of 10 mm.
Spirit level
► If the device cannot be aligned horizontally with the screws screwed
Air bubble
Function
min
min. 0,5 m
mi
min. 0,5 m
out as far as possible, the setup location is unsuitable. Set the
device up in a suitable location.
English
min. 0,2 m
min. 0,5 m
Starting up the device • 93