Element
STANDARD 2
FAN SPEED
LEAK COUNTS
Element
LAST MEAS CHANNEL Last channel measured.
LAST MEAS TIME
LAST ABS
LAST CONC
NEXT MEAS TIME
DARK
REF
SAMPLE
DARK STD DEV
REF STD DEV
SAMPLE STD DEV
SAMPLE VOLUME
REAGENT 1
REAGENT 2
REAGENT 3
View instrument information
1. Select INSTRUMENT INFORMATION.
2. Select an option.
Option
ANALYZER INFO
MODULE INFO (option is available when a
module is installed)
18 English
Table 2 Analyzer data (continued)
Definition
Level of standard solution remaining
The speed of the ventilation fan.
Indication of possible fluid leaks (range 0 to 1023). A count of over 511 indicates a fluid
leak
Table 3 Measurement data
Definition
Time of last measurement.
Last absorbance reading.
Concentration of last measurement.
Time the next measurement will be taken.
Number of A/D counts measured when the LED is turned off.
Reference A/D count used to compensate for natural color and turbidity.
A/D counts measurement (after color development) used to determine the
concentration of the sample.
Standard deviation of dark counts out of 6 readings.
Standard deviation of reference counts out of 6 readings.
Standard deviation of sample counts out of 6 readings.
Total sample flush volume through the colorimeter for the measurement cycle.
Calculated reagent delivery time to the sample based on temperature, pressure and
viscosity.
Calculated reagent delivery time to the sample based on temperature, pressure and
viscosity.
Calculated reagent delivery time to the sample based on temperature, pressure and
viscosity.
Description
Shows the software information and serial number.
Shows the connected modules with software information
and serial number.