Testing Instructions
for (Re-) Calibration Checks
A) Function testing
A function test can only be carried out on a
person or with a simulator.
B) Testing of Density of Pressure Circuit and
Divergence from Pressure Indication
1. Remove batteries.
2. Hold
-button pressed and at the same time
insert batteries.
3. Release
-button. The unit now is ready for
testing. The displays of SYS and DIA now
show the actual pressure.
4. Prior to testing the density, open the battery
compartment and take the T-connector off the
unit. Wrap the cuff around a fixed cylinder and
connect it with the T-connector (without defla-
tion valve) to the pump ball and the unit. Mind
the setting time for cuff: at least 30 seconds.
Accomplish testing in the usual manner.
5. Prior to testing the divergence of the pressure
indication, open the battery compartment and
take the T-connector off the unit. Connect the
unit with the normal unit and a pump ball.
Accomplish testing in the usal manner.
6. Switch off unit.
7. Re-connect the cuff and the pumpball with the
T-connector.
C) Placing of Safety Stamps
For complete safety cover up the drill hole for the
base screw.
Alternatively, connect upper housing and lower
housing with a safety stamp placed directly on
one of the housing edges.