6. Maintenance Intervals
Inspection and maintenance
Operational check
Coarse dust filter
Fine filter
Microfilter
Patient bacteria filter
Inhalation equipment (e.g. nasal canula) Exchange according to instructions of
and oxygen extension hose
Note! All maintenance intervals are guidelines. In different cases these can be shorter due to changing
conditions regarding temperature, humidity, dust etc. The filter service life is 20% less in devices
with 6 l capacity.
7. Specifications
Ambient temperature running:
Dimensions:
Weight:
Line voltage:
Nominal current:
Flow:
Oxygen concentration
at 20°C, 1013 hPa
50% rel. humidity
Sound pressure level
Output pressure
All manuals and user guides at all-guides.com
Once a year or after max. 2800 working
hours by the service technician responsible
Every six month by service technician responsible
Clean filter fleece once a week. If blocked,
exchange by the service technician responsible.
Exchange every six month or after max. 2000 working hours
by the service technician responsible.
Exchange every six month or after max. 2000 working hours
by the service technician responsible
Exchange during inspection and maintenance by the
service technician responsible
the attending physician
Model (RC 5)
10 to 40° C
55 x 21.5 x 55 cm
approx. 21 kg
230 V +/-10%
50 Hz
2.0 A
0.1 - 5 l/min
0.1 - 3 l/min
95 +/- 3 Vol.% O2
-4 l/min
90 +/- 3 Vol.% O2
- 5 l/min
82 +/- 3 Vol.% O2
48 dB(A)
300-350 mbar
Model (RC 5) S
10 to 40° C
55 x 21.5 x 55 cm
approx. 21 kg
230 V +/-10%
50 Hz
2.0 A
0.1 - 6 l/min
0.1 - 4 l/min
95 +/- 3 Vol.% O2
-5 l/min
88 +/- 3 Vol.% O2
- 6 l/min
80 +/- 3 Vol.% O2
49 dB(A)
300-350 mbar
17