Symptom
Compressor off
Display: Temperature
Interval sound
Compressor off Display:
Pressure too low
Interval sound
Compressor off Display:
Pressure too high
Interval sound
Compressor off Display:
Column-valve
Device is running Display:
Check Head
Interval sound
Device is running
Display: Ions not OK
Interval sound
5. Cleaning
Turn off the device and pull the plug from mains
before cleaning.
The external box may be cleaned with custo-
mary cleaning agents. Do not, however, use any
caustic, scratching or acidic cleaning agents.
Do not bring water into the device's interior.
Daily
Rinse the nasal canula with clear, warm water.
When an autoclavable effervescent humidifyer
is used it should be cleaned, respectively autocla-
ved, according to the instructions given by the
producer of that autoclave effervescent humidify-
All manuals and user guides at all-guides.com
Cause
Device became too hot
due to wrong placing
Device became too hot because
coarse dust filter is blocked
Device defect
Coarse dust filter blocked
Intake filter blocked
Device defect
Device defect
Hose connection
valve-column not tight
Ionization head not screwd
on correctly
Ionization head is wet
Device defect
Remedy
Let device cool off for 30 min.,
place correctly, turn on again
Clean filter fleece and turn on again
Call service
Clean filter fleece and turn on again
Call service
Call service
Call service
Call service
Screw on ionization head correctly
Dry ionization head
Call service
er. Non-observance of the instructions can lead to
a contamination of the effervescent humidifyer.
Option I
Clean Ionization Head
Unscrew ionization head and rinse with clear,
warm water. Boil out once a week for 10 to 15 mi-
nutes clean in dishwasher at maximum tempera-
ture. Leave to dry afterwards. The mask and the
ionization head can be treated with the desinfec-
tion spray offered. Use only dry ionization heads.
16