I N T E L E C T ® M O B I L E 2 Q U I C K S TA RT G U I D E
APPENDIX
IFC Premodulated (Traditional 2 Pole)
Advice on size and type of electrodes to be used is given in device GUI "treatment
guidelines" feature
Premodulated Current is a medium frequency waveform. Current
comes out of one channel (two electrodes). The current intensity is
modulated: it increases and decreases at a regular frequency (the
Amplitude Modulation Frequency).
Output Mode
Available on Channel
Treatment Time (STIM)
Treatment Time (COMBO)
Mode Selection
Output Intensity
Beat Fixed (Sweep Off)
Cycle Time
Frequency
Sweep Low Beat Frequency
Sweep High Beat Frequency
IRMS
DC component
22
Electrodes
1, 2
1-60 Min
1-30 Min
CC or CV
0-100 mA (CC) 0-100 V (CV)) Carrier
1-200 Hz
Continuous or User Defined
2,000-10,000 Hz
1- 199 Hz
2-200 Hz
0-55mA
No
High Voltage Pulsed Current (HVPC)
Advice on size and type of electrodes to be used is given in device GUI "treatment
guidelines" feature
The High Voltage Pulsed Current (HVPC) has a very brief pulse
duration characterized by two distinct peaks delivered at high
voltage. The waveform is monophasic (current flows in one
direction only). The high voltage causes a decreased skin resistance
making the current comfortable and easy to tolerate.
Output Mode
Available on Channels
Treatment Time (Stim)
Treatment Time (Combo)
Mode Selection
Output Intensity
Cycle Time
Display
Frequency
Polarity
Ramp
Sweep time
Sweep High Frequency
Sweep Low Frequency
IRMS
DC component
Electrodes
1, 2
1-60 Min
1-30 Min
CV
0-500 V (CV)
Continuous or User Defined
Volts
1-200 pps
Positive or Negative
0.5-5 sec
14sec
2-200 pps
1-199 pps
0-45mA
0 - 1.5mA