14.5 Use of the microscope (B-383FL)
1. Turn on the power supply for the mercury bulb and wait 5
minutes for the arc to stabilize.
2. Move the filter selector ① to one of the 2 available positions
until the click stop. (Fig. 49).
3. The microscope has a 3-position filter holder. The leftmost
position allocates the B filter, the central position is empty
for transmitted light observation and the rightmost position
allocates the G filter.
14.6 Use of the microscope (B-383LD)
1. Turn on the fluorescence LED, by switching on "II" the main
switch placed on the back side of the frame. (Fig. 50)
2. Move the filter slider
FILTER
EXCITATION
CUBE
FILTER
B
475/30 nm
G
530/40 nm
all the way in. (Fig. 51).
DICHROIC
MIRROR
505 nm
570 nm
EMISSION
FILTER
FITC: fluorescent antibodies
515LP nm
Achridine orange: DNA, RNA
Auramine
Rhodamine, TRITC: fluorescent antibodies
590LP nm
Propidium iodide: DNA, RNA
RFP
Page 32
APPLICATIONS
F ig. 49
F
ig. 49
F ig. 50
F
ig. 50
F ig. 51
F
ig. 51