ONLY FOR B-383FL
Disconnect all electrical cables before installing or replacing the bulb.
The bulb has an anode and a cathode of different sizes. Respect the polarity during assembly, respecting the
bulb dimensions.
Do not touch the bulb with bare hands to leave no traces of grease on the bulb. If this happens, clean the bulb
with a soft cloth before turning on the lamp.
The bulb has an average life of about 200-250 hours: a time counter and a voltage indicator are shown on the
bulb power supply. Replace the bulb when the hour count exceeds 250 or if the voltage drops below 4.5A.
During use, the bulb, the lamp housing and the surrounding environment become hot.
Before replacing the bulb, switch off the power supply, disconnect all cables and wait for the bulb and the lamp
housing to cool.
After switching on the bulb, wait at least 10-15 minutes before switching it off.
After switching off the bulb, wait for 5-10 minutes before switching it on again so that the mercury vapors have
time to condense.
The bulb emits ultraviolet radiation that could be harmful to eyes and skin.
14.3 HBO bulb assembling (B-383FL)
1. Open the lamp housing using the door lock screw ① and
remove the lamp holder. (Fig. 37)
2. Remove the plastic block ② from the lamp holder (or the
exhausted lamp in case of replacement) by loosening the two
locking screws ③. (Fig. 38)
3. Insert the mercury bulb ④ (respect the polarity of the bulb),
tighten the locking screws and refit the lamp holder inside the
lamp housing. (Fig. 39)