ONLY FOR B-383FL
•
Disconnect all electrical cables before installing or replacing the bulb.
•
The bulb has an anode and a cathode of different sizes. Respect the polarity during assembly, respecting the
bulb dimensions.
•
Do not touch the bulb with bare hands to leave no traces of grease on the bulb. If this happens, clean the bulb
with a soft cloth before turning on the lamp.
•
The bulb has an average life of about 200-250 hours: a time counter and a voltage indicator are shown on the
bulb power supply. Replace the bulb when the hour count exceeds 250 or if the voltage drops below 4.5A.
•
During use, the bulb, the lamp housing and the surrounding environment become hot.
•
Before replacing the bulb, switch off the power supply, disconnect all cables and wait for the bulb and the lamp
housing to cool.
•
After switching on the bulb, wait at least 10-15 minutes before switching it off.
•
After switching off the bulb, wait for 5-10 minutes before switching it on again so that the mercury vapors have
time to condense.
•
The bulb emits ultraviolet radiation that could be harmful to eyes and skin.
14.3 HBO bulb assembling (B-383FL)
1. Open the lamp housing using the door lock screw ① and
remove the lamp holder. (Fig. 37)
2. Remove the plastic block ② from the lamp holder (or the
exhausted lamp in case of replacement) by loosening the two
locking screws ③. (Fig. 38)
3. Insert the mercury bulb ④ (respect the polarity of the bulb),
tighten the locking screws and refit the lamp holder inside the
lamp housing. (Fig. 39)
Page 28
①
F ig. 37
F
ig. 37
③
②
③
F ig. 38
F
ig. 38
④
F ig. 39
F
ig. 39