TB3 – Conventional (analog) Sensors
pH
Term. Function
1
Glass
2
3
Reference**
4
Solution GND / Shield** green / yellow and blue
5
6
RTD ret / GND
7
RTD sense
8
RTD
9
+5V
TB3 – Conventional (analog) Sensors
Cond Ind
Term. Color. InPro 7250 ST / PFA Color. InPro 7250 HT
1
Coax inner / transparent
2
red
3
green / yellow
4
brown, white
5
blue
6
white
7
grey
8
green
9
TB3 – Conventional (analog) Sensors
Dissolved carbon dioxide
Term. Color*
1
Coax inner / transparent
2
3**
Coax shield / red
4**
green / yellow
5
6
white
7
8
green
9
Color*
tansparent
red
white
green
Coax inner / transparent receive hi
yellow
green / yellow
violet
black
white
grey
green
Function
Glass
not used
Reference
GND / Shield
not used
RTD ret / GND
not used
RTD
+5V
Redox (ORP)
Function
Color
Platinum
transparent***
Reference
Shield
Function
receive lo
shield / GND
send lo
send hi
RTD ret / GND
RTD sense
RTD
not used
Terminal 4 and 6 are inter-
nally connected, either ter-
minal can be used to con-
nect a wire.
* Grey wire not used.
** Jumper 3 to 4 has to be
installed.
Terminal 4 and 6 are inter-
nally connected, either ter-
minal can be used to con-
nect a wire.
*
Grey wire not used.
** For pH without solution
ground, install jumper
3 to 4.
*** Remove black coating.
Terminal 4 and 6 are inter-
nally connected, either ter-
minal can be used to con-
nect a wire.