10. Specifications
Weight
Dimensions
Cuff circumference
applicable to the
monitor
Memory
Contents
Protection against
electric shock
Applied part
Note
•These specifications are subject to change without notice.
•This monitor complies with the requirements of ISO 81060-2:2013.
In the clinical validation study, K5 was used on 85 subjects for
determination of diastolic blood pressure.
•This monitor has not been validated for use on pregnant patients.
•IP classification is degrees of protection provided by enclosures in
accordance with IEC 60529. This monitor and AC adapter are
protected against solid foreign objects of 0.5 inch (12.5 mm)
diameter and greater such as a finger. The AC adapter is protected
against vertically falling water drops which may cause issues during
a normal operation.
50
Monitor: Approximately 16.6 oz. (470 g) not
including batteries
Arm cuff: Approximately 6 oz. (170 g)
Monitor: Approximately 7.5" (w) × 3.1" (h) ×
4.1" (l) (191 mm × 79 mm × 105 mm)
Arm cuff: Approximately 5.7" × 23.4" (air
tube: 29.5") (145 mm × 594 mm (air tube:
750 mm))
7" to 17" (170 to 420 mm) (included arm cuff:
9" to 17" (220 to 420 mm))
Stores up to 100 readings per user
Monitor, Wide-Range D-Ring Cuff, AC
adapter (HEM-ADPTW5), 4 "AA" batteries,
storage case, instruction manual, quick start
guide
Internally powered ME equipment (When
using only batteries)
Class II ME equipment (AC adapter)
Type BF (arm cuff)