6. What do I do if...
... no weight display comes on?
– Is the scale switched on?
– Check the mains connection and the connection of the battery pack.
– Switch the scale off for a few seconds. This can be done by opening the plug connect-
ed to the rechargeable battery block and unplugging the mains supply from the scale.
0. 0
...
does not appear before weighing?
– Switch off the scale with the Start key and start the scale again – there must not be
any load on the scale – and only its feet should be in contact with the floor.
... one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
... the display shows
– Battery voltage is dropping. Charge the rechargeable batteries in the next few days.
...
batt
appears in the display?
– The rechargeable battery is so severely discharged that weighing is no longer possible.
Recharge the battery.
...
St0p
appears in the display?
– Maximum load has been exceeded.
... the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
0. 0
and wait until
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
E
... the display
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, disconnect the power supply by briefly removing the batteries. If
this measure is equally unsuccessful, inform the Service department.
7. Servicing
Your seca scale leaves the factory with an accuracy greater than ± 0.1 %. To ensure that
this accuracy is maintained, the product must be carefully set up and regularly maintained.
Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please
ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for
further information.