8.3
Motorised version
1. Assemble the stage in the same way as the manual version.
Check the perfect alignment of the rear part of the stage with
the rear arm of the frame.
A non perfect alignment could lead to an incorrect functioning
of the system. (Fig. 16)
2. Connect the cable ① from the stage to the frame and tight
the locking screws of the connectors ②. (Fig. 17)
3. Connect the provided cables: ③ 12V power supply for the
motorised parts; ④ 6V microscope power supply; ⑤ serial
cable; ⑥ PS/2 mouse. (Fig. 18)
Connect power cables as the last step.