English
VADOplex
Information on potential electromagnetic interference and advice on how to
avoid or minimize such interference
Guidance and manufacturer's declaration – electromagnetic emission
The model VADOplex is intended for use in the electromagnetic environment specified below.
The customer or the user of the model VADOplex should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment
and the model VADOplex
The model VADOplex is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the model VADOplex can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the model VADOplex as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
Separation distance according to frequency of transmitter
power of
transmitter
150 kHz
W
to 80 MHz
d = 1,2
0,01
0,12
0,1
0,38
1
1,2
10
3,8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
48
Electromagnetic
environment – guidance
The model VADOplex uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The model VADOplex is suitable
for use in all establishments,
including domestic establishments
and those directly connected
to the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
m
80 MHz
800 MHz
to 800 MHz
to 2,5 GHz
d = 1,2
d = 2,3
0,12
0,23
0,38
0,73
1,2
2,3
3,8
7,3
12
23
Information on potential electromagnetic interference and advice on how to
avoid or minimize such interference
Guidance and manufacturer's declaration – electromagnetic immunity
The model VADOplex is intended for use in the electromagnetic environment specified below. The customer
or the user of the model VADOplex should assure that it is used in such an environment. The device
complies with the requirements (electromagnetic immunity) listed in Section 8.9 of IEC 60601-1-2:2014.
Immunity test
IEC60601-1-2 test
level
Electrostatic
± 8 kV contact
discharge (ESD)
± 2 kV, ± 4 kV,
IEC 61000-4-2
± 8 kV, ± 15 kV air
Power frequency
30 A/m
(50/60Hz) magnetic
field
IEC 61000-4-8
Radiated RF EM
IEC 60601-1-2:2014
fields: Proximity
table 9
fields from RF
wireless
communications
equipment
IEC 61000-4-3
Electrical fast
±2 kV
transient/burst
100 kHz Repetition
IEC 61000-4-4
Frequency
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common
mode
Compliance level
Electromagnetic environment –
guidance
± 8 kV contact
Floors should be wood,
± 2 kV, ± 4 kV,
concrete or ceramic tile. If floors
± 8 kV, ± 15 kV air
are covered with synthetic
material, the relative humidity
should be at least 30 %.
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location of a professional
healthcare facility environment
and Home healthcare
environment.
IEC 60601-1-2:2014
table 9
±2 kV
Mains power quality should be
100 kHz Repetition
that of a professional healthcare
Frequency
facility environment and Home
healthcare environment.
±1 kV differential
Mains power quality should be
mode
that of a professional healthcare
±2 kV common
facility environment and Home
mode
healthcare environment.
49