English
VADOplex
Useful details
Technical data
MDD class
IIa (93/42/EEC, Annex IX, Rule 9)
Protection class
I, Type B
CE certification
No. CE 0123
Rating
Input:
230 V
Power supply:
Manufactured by
M+R Multitronik GmbH
Power supply model:
ACS60M-24
Fuses:
T2AH250V
DIMDI No.
10-969 in accordance with UMDNS
Classification
UMDNS
compression device intermittent
Code of the authority
responsible DE CA 57 (Upper Bavaria)
Article-No.
VXSY-07-1
This device is intended for continuous use. EMC tested in accordance with IEC 60601-1
and IEC 60601-1-2.
For information about potential electromagnetic interference and advice on how to avoid
or minimise such interference, please refer to the attachment of this user manual.
Environmental conditions for transport/storage
Temperature range: -25° C to 70° C
●
●
Relative humidity: up to 95 % without condensation
Ambient pressure: 700 hPa to 1.060 hPa
●
Environmental conditions during operation
Temperature range: 10° C to 40° C
●
●
Relative humidity: up to 95 % without condensation
Ambient pressure: 700 hPa to 1.060 hPa
●
Device dimensions
Height:
23 cm, overall
Depth without handle:
15.5 cm
Width:
35 cm
Weight:
4.0 kg
Warranty
The warranty complies with statutory regulations. The terms and conditions of business of OPED
GmbH also apply, where permissible.
46
0123
, 50/60 Hz, 0,4 A
Useful details
Conversion table
mmHg
KPa
mmHg
60
8,0
110
70
9,3
120
80
10,7
130
90
12,0
140
100
13,3
150
Risk of falling
Risk of explosion
Before opening the device, unplug the
power supply.
The device meets all essential requirements
of the applicable EU guidelines.
0123
Type B applied part.
Observe accompanying documents!
The product and its packaging may not be
disposed of as household waste, but has to be
sent back to the manufacturer for disposal.
The manufacturer commits to the legally
compliant separation and collection of electrical
and electronic equipment based on the
requirements of ElektroG and the WEEE directive.
KPa
mmHg
KPa
14,7
160
21,3
16,0
170
22,7
17,3
180
24,0
18,7
190
25,3
20,0
200
26,7
Manufacturer
Date of manufacture
40°C
Temperature range
10°C
Data-Matrix
Safety Note
Single patient use only: The pads are to be
used for one patient only and may neither be
used for several patients nor refurbished.
47