VADOplex
1. Mode: Program settings
In the selection level
, confirm the setting with the
11
Select a different mode and press the
-button.
Program 1
8
Program 2
Foot
Foot for application
in the case of arterial
circulatory disorders/Calf
• 20 sec. cycle time
• 50 sec. cycle time
• 130 mmHg/
• 130 mmHg/
pressure impulse
pressure impulse
duration 1 seconds
duration 3 seconds
Press the
-button to return to the selection level; press the
application level. An overview of the device settings is found at the front of the instructions
for use.
17
Note: If 3 seconds application time is set up for both sides, the interval between
the impulses will raise automatically from 20 seconds to 50 seconds! (due to slower
venous refill time with arterial diseases)
Note: If there is a power outage, the device has to be switched on again (see Starting-
up the Device). In this case, the controller automatically resets to the default factory
settings. Any changes made to the pressure settings/programme are lost and have to be
reset.
2. Beeper: Volume control for beep signals
In the selection level
, confirm the setting with the
12
can change the volume here. Choose a volume from 1 to 8 (quiet to loud).
with volume 8)
Press the
-/
-button on the right to decrease or increase the volume. Press the
-button to confirm the input. The input is indicated by brief flashing the mmHg/secs.
Press the
-button to return to the selection level; press the
application level. An overview of the device settings is found at the front of the instructions
for use.
17
3. Timer: Application duration setting
In the selection level
, confirm the setting with the
14
The application duration for the patient can be configured in the Timer menu. The selection
range is 0-15h.
38
-button to enter the submenu.
-/
-button on the right. Confirm with the
9
Progrmm 3
10
Hand
• 20 sec. cycle time
• 80 mmHg/
pressure impulse
duration 1 second
-button again to go to the
9
-button to enter the submenu. You
(
, illustrated
13
-button again to go to the
-button to enter the submenu.
Press the
-/
-button on the right to configure the desired application time ("time" flashes
on the display). Press the
-button to confirm the input. The input is indicated by brief
flashing of the mmHg/secs.
Press the
-button to return to the selection level; press the
application level. An overview of the device settings is found at the front of the instructions
for use.
17
When the timer is active, the symbol (manikin) and 'hrs' (time in hours) flash on the centre of
the display.
(
, illustrated with 3 of 5 hours elapsed)
15
About every 12 seconds the
Resetting the patient hours
Switch on the power switch.
After the device is switched on, the power switch is lit up
4
green. A welcome messages is shown on the display.
4-3-2-1-0), shown at the top left on the display.
tient hours to 0, the
-button on the right has to be pressed and held during the 5-second
countdown (5-4-3-2-1-0) while the device is powered up. An overview for putting the device
into operation is found at the front of the instructions for use.
-button again to go to the
display switches to the respective selected program.
A 5-second countdown follows (5-
6
(
, illustrated at 3 seconds)
To reset the pa-
7
16
39