Display Information; Device Settings - OPED VADOplex Mode D'emploi

Masquer les pouces Voir aussi pour VADOplex:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

English
VADOplex

Display information

D
Application time in seconds, left side
Pressure setting left side
O
OK: displayed as soon as set pressure is reached
Pressure setting right side
Q
Application time in seconds, right side
Display between pressure pulses
S
Display during pressure pulses
T
Application
To make the device ready for use, please read at first the safety notes on page 28 and 29
before you follow the installation steps like below.
Connect the connecting hose to the pad hose. To do so, insert the metal connector on the
connecting hose into the pad hose.
Attach the VADOplex Pads and make sure that they are positioned properly
(see user manual supplied with pads).
Single patient use only: The pads are to be used for one patient only and may neither
be used for several patients nor refurbished.
-button on the left side of the control panel is for the left application.
The
Simultaneously pressing both
-/
(see Device Settings)
The
-button on the right side of the control panel is for the right application.
The
and
-/
-buttons on the right side of the control panel are for navigation in the
Device Settings menu.
A different screen can be displayed depending on the individual start-up settings
screen
,
,
).
8
9
10
Program 1
8
Program 2
Foot
Foot for application
in the case of arterial
circulatory disorders/Calf
• 20 sec. cycle time
• 50 sec. cycle time
• 130 mmHg/
• 130 mmHg/
pressure impulse
pressure impulse
duration 1 seconds
duration 3 seconds
The factory default start-up setting is 130 mmHg with a pressure pulse duration of 1 second.
Always chose the modes by pressing the
the
-button on the right or left. Press the
(mmHg) can be changed by pushing the
After the application has ended, press the
the pad and switch the controller off using the power switch
cable from the socket.
36
N
P
R
-buttons navigates you to the Device Settings menu
(for example
9
Progrmm 3
10
Hand
• 20 sec. cycle time
• 80 mmHg/
pressure impulse
duration 1 second
-button. If you want to change the settings, press
-button to start the treatment. The pressure
- or
-button.
-button. The programme is stopped. Remove
. Be sure to remove the power
E
Pause mode
If you want to interrupt the Application mode, press the
again, you can continue the application mode.
5
Rest mode
The Rest mode always sets in when the controller is in operation, but is not actively working.
You can exit the Rest mode by pressing any button.
Timer mode
After the previously set application time has expired (timer), the device enters rest mode.
Note: If 3 seconds application time is set up for both sides, the interval between the
impulses will raise automatically from 20 seconds to 50 seconds! (due to slower venous
refill time with arterial diseases)
9
Switching off the device
After the application has ended, press the
the pad and switch the controller off using the power switch
power cable from the socket.
Note: If there is a power outage, the device has to be switched on again (see Starting-
up the Device). In this case, the controller automatically resets to the default factory
settings. Any changes made to the pressure settings/programme are lost and have to be
reset.

Device settings

To enter the device settings menu, both
time during the 5-second countdown (5-4-3-2-1-0) while starting up the device. An overview
for putting the device into operation is found at the front of the instructions for use.
Illustration at 3 seconds)
Three different program settings can be pre-selected in the device settings menu:
1. Mode (program)
11
2. Beeper (volume)
12
3. Timer (application duration setting)
Switch between the three different settings by pressing the
Press the
- button to confirm the setting and go to the respective submenu. Leave the
submenu by pressing the
-button. This takes you back to the selection level. Press the
-button again to return to the application level. An overview of the device settings is
found at the front of the instructions for use.
-button. By pressing this button
-button. The programme is stopped. Remove
. Be sure to unplug the
E
-buttons have to be pressed briefly at the same
14
-/
-button on the right
17
(
,
16
.
17
37