Agilent TwisTorr 305-IC Manuel De L'utilisateur page 211

Table des Matières

Publicité

Technical Information
Compliance with:
Power cable
USB communication
When the TwisTorr 305-IC has been stored at a temperature less than 5°C, wait
NOTE
until the system has reached the above mentioned temperature before switch the
TwisTorr 305-IC on.
Figure 44 Maximum allowed relative humidity as a function of the air temperature for each
Agilent TwisTorr 305-IC User Manual
**
cooling-water temperature
EN 61010-1
EN 61326-1
EN 1012-2
EN 12100
EN 50581
Machinery Directive 2006/42/EC
Electromagnetic Compatibility Directive
2014/30/EU
Directive 2011/65/EU
The admitted power supply voltage value
(measured at the electronics' input connector) is
24V +/- 10%.
Please size the supply cables in order to always
ensure that the minimum voltage value at the input
connector is respected. For instance: the maximum
length that is allowed for a typical AWG20 cable is
equal to 3m. (Typ. resistance value for AWG20
cable: 33,3mOhm/m; in case of a single pole wire
with a length equal to 6m and with the maximum
current supplied to the controller (7.5A), the voltage
drop on the cable will be equal to 1.5V).
as per USB 1.0