Instructions for Use

Stopping the TwisTorr 305-IC

To stop the TwisTorr 305-IC, not in emergency, it is necessary to:
a) if the TwisTorr 305-IC is configured in REMOTE MODE, and the INTERLOCK is
set to CONTINUOUS MODE, then open the "START" or the "INTERLOCK" signals;
b) if the TwisTorr 305-IC is configured in REMOTE MODE, and the INTERLOCK is
set to IMPULSE MODE, then open the "START" signal;
c) if the TwisTorr 305-IC is configured in SERIAL Mode, then send a STOP
command to the TwisTorr 305-IC.

Emergency stop

Emergency stop can be achieved by removing the power supply.
This mode is to be used only in an emergency. The use of this stop mode
WARNING
could lead to faults and/or damages at the pump. Improper use of this
stop mode could void the warranty.