Instructions for Use

TwisTorr 305-IC Mounting

Failure to comply with these installation instructions could result in the
WARNING
pump detaching from the system in the event of a rotor failure, which
could cause property damage or serious injury or death.
The TwisTorr 305-IC can be installed in any position.
Fix the TwisTorr 305-IC in a stable position, mounting the inlet flange of the
turbopump to the system counter-flange, with a connection capable of
withstanding a torque of 1000 Nm around its axis.
The turbopump with ConFlat and ISO K inlet flange must be fixed to the vacuum
chamber by means of the appropriate Agilent hardware.
For further details see the "Installation of ConFlat and ISO K" section.
Power supply cord: The correct cable for electrical wiring is a three wires
(Ph+N+Earth) cable. The wire section has to be at least 0.75 mm
The TwisTorr 305-IC cannot be fixed by means of its base or through other
CAUTION
flanges that are not inlet.
Agilent TwisTorr 305-IC User Manual
2
(AWG18).