Instructions for Use

Connection of the Fore-Vacuum Pump

The foreline connection should be made using the KF 16 NW flange provided on the
TwisTorr 305-IC. The connection to the foreline pump can be made by a vacuum
rated hose or pipe. If a rigid pipe is used, any vibration generated by the mechanical
pump must be damped through the use of bellows.
The TwisTorr 305-IC is characterized by its high compression ratio also for oil
NOTE
vapors. When using a mechanical oil-sealed pump, it is advisable to install a suitable
trap between the turbopump and the fore-vacuum pump in order to prevent back
streaming oil or particulate from dry pumps.